Improved cardiovascular flow quantification with time-resolved volumetric phase-contrast MRI

Pediatr Radiol. 2011 Jun;41(6):711-20. doi: 10.1007/s00247-010-1932-z. Epub 2011 Jan 11.

Abstract

Background: Cardiovascular flow is commonly assessed with two-dimensional, phase-contrast MRI (2-D PC-MRI). However, scan prescription and acquisition over multiple planes is lengthy, often requires direct physician oversight and has inconsistent results. Time-resolved volumetric PC-MRI (4-D flow) may address these limitations.

Objective: We assess the degree of agreement and internal consistency between 2-D and 4-D flow quantification in our clinical population.

Materials and methods: Software enabling flow calculation from 4-D flow was developed in Java. With IRB approval and HIPAA compliance, 18 consecutive patients without shunts were identified who underwent both (1) conventional 2-D PC-MRI of the aorta and main pulmonary artery and (2) 4-D flow imaging. Aortic and pulmonary flow rates were assessed with both techniques.

Results: Both methods showed general agreement in flow rates (ρ: 0.87-0.90). Systemic and pulmonary arterial flow rates were well-correlated (ρ: 4-D 0.98-0.99, 2-D 0.93), but more closely matched with 4-D (P < 0.05, Brown-Forsythe). Pulmonary flow rates were lower than systemic rates for 2-D (P < 0.05, two-sample t-test). In a sub-analysis of patients without pulmonary or aortic regurgitation, 2-D showed improved correlation of flow rates while 4-D phase-contrast remained tightly correlated (ρ: 4-D 0.99-1.00, 2-D 0.99).

Conclusion: 4-D PC-MRI demonstrates greater consistency than conventional 2-D PC-MRI for flow quantification.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Algorithms*
  • Cardiac-Gated Imaging Techniques / methods*
  • Child
  • Child, Preschool
  • Coronary Circulation*
  • Female
  • Humans
  • Image Enhancement / methods*
  • Magnetic Resonance Angiography / methods*
  • Magnetic Resonance Imaging, Cine / methods*
  • Male
  • Pulmonary Circulation*
  • Reproducibility of Results
  • Sensitivity and Specificity