Approximate truncation robust computed tomography--ATRACT

Phys Med Biol. 2013 Sep 7;58(17):6133-48. doi: 10.1088/0031-9155/58/17/6133. Epub 2013 Aug 14.

Abstract

We present an approximate truncation robust algorithm to compute tomographic images (ATRACT). This algorithm targets at reconstructing volumetric images from cone-beam projections in scenarios where these projections are highly truncated in each dimension. It thus facilitates reconstructions of small subvolumes of interest, without involving prior knowledge about the object. Our method is readily applicable to medical C-arm imaging, where it may contribute to new clinical workflows together with a considerable reduction of x-ray dose. We give a detailed derivation of ATRACT that starts from the conventional Feldkamp filtered-backprojection algorithm and that involves, as one component, a novel original formula for the inversion of the two-dimensional Radon transform. Discretization and numerical implementation are discussed and reconstruction results from both, simulated projections and first clinical data sets are presented.

MeSH terms

  • Algorithms*
  • Cone-Beam Computed Tomography / methods*
  • Image Processing, Computer-Assisted / methods*
  • Phantoms, Imaging
  • Radiography, Thoracic