Evaluation of occurring complications after flow diverter treatment of elastase-induced aneurysm in rabbits using micro-CT and MRI at 9.4 T

Neuroradiology. 2016 Oct;58(10):987-996. doi: 10.1007/s00234-016-1730-x. Epub 2016 Jul 25.

Abstract

Introduction: Flow diverters are increasingly being used to treat intracranial aneurysms. This study evaluates occurring complications of flow-diverting devices in the treatment of experimental aneurysms, involving the use of micro-CT and small animal MRI at 9.4 T, in correlation to angiographic and histological findings.

Methods: We previously published two preclinical studies, in which we assessed two different flow diverters in the treatment of elastase-induced aneurysms. Devices have been implanted across the aneurysm neck as well as in the abdominal aorta. From these studies, a total of 65 devices (prototype FD (n = 30) and Derivo embolization device (n = 35)) additionally underwent micro-CT and MRI after angiographic follow-up and before being histologically examined.

Results: The different architectures of both devices were precisely comparable due to high-resolution micro-CT imaging. Micro-CT revealed wire fractures in nine cases (30 %) only with the prototype FD. In three cases (10 %), severe wire fractures correlated with an in-stent stenosis due to intimal hyperplasia. Other complications, like distal stent occlusions and post-stent stenosis, were seen in both groups and verified with both imaging techniques. Osseous metaplasia were correlated to calcifications seen with micro-CT. MRI enabled visualization of the position of the implanted devices relative to the aneurysm and revealed incomplete aneurysm neck coverage with the prototype FD in two cases (6.7 %).

Conclusion: Micro-CT and 9.4-T MRI are valid to discover and understand occurring complications of flow diverters in the preclinical phase and can serve as evaluation tools to minimize complication rates of endovascular devices in the future.

Keywords: 9.4-T MRI; Aneurysm; Animal studies; Flow diverter; Micro-CT.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Animals
  • Cerebral Angiography / methods
  • Cerebral Revascularization / adverse effects*
  • Cerebral Revascularization / instrumentation*
  • Cerebral Revascularization / methods
  • Computed Tomography Angiography / methods
  • Equipment Design
  • Equipment Failure Analysis
  • Female
  • Graft Occlusion, Vascular / diagnostic imaging
  • Graft Occlusion, Vascular / etiology*
  • Intracranial Aneurysm / chemically induced
  • Intracranial Aneurysm / diagnostic imaging*
  • Intracranial Aneurysm / therapy*
  • Magnetic Resonance Angiography / methods*
  • Pancreatic Elastase
  • Rabbits
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Treatment Outcome
  • X-Ray Microtomography / methods*

Substances

  • Pancreatic Elastase