Interchangeable neck shape-specific coils for a clinically realizable anterior neck phased array system

Magn Reson Med. 2017 Dec;78(6):2460-2468. doi: 10.1002/mrm.26632. Epub 2017 Feb 10.

Abstract

Purpose: To demonstrate the interchangeable neck shape-specific (NSS) coil concept that supplements standard commercial spine and head/neck coils to provide simultaneous high-resolution (hi-res) head/neck imaging with high signal-to-noise ratio (SNR).

Methods: Two NSS coils were constructed on formers designed to fit two different neck shapes. A 7-channel (7ch) ladder array was constructed on a medium neck former, and a 9-channel (9ch) ladder array was constructed on large neck former. Both coils were interchangeable with the same preamp housing.

Results: The 7ch and 9ch coils demonstrate SNR gains of approximately 4 times and 3 times over the Siemens 20-channel head/neck coil in the carotid arteries of our volunteers, respectively. Coupling between the Siemens 32-channel spine coil, Siemens 20-channel head/neck coil, and the NSS coils was negligible, allowing for simultaneous hi-res head/neck imaging with high SNR.

Conclusions: This study demonstrates that supplementing existing commercial spine and head/neck coils with an NSS coil allows uniform simultaneous hi-res imaging with high SNR in the anterior neck, while maintaining SNR of the commercial coil in the head and posterior neck. Magn Reson Med 78:2460-2468, 2017. © 2017 International Society for Magnetic Resonance in Medicine.

Keywords: RF coil; cervical; magnetic resonance imaging; neck; phased array; signal-to-noise ratio.

MeSH terms

  • Carotid Arteries / diagnostic imaging
  • Carotid Artery Diseases / diagnostic imaging*
  • Chin / diagnostic imaging
  • Computer Simulation
  • Equipment Design
  • Healthy Volunteers
  • Humans
  • Image Processing, Computer-Assisted
  • Magnetic Resonance Imaging / instrumentation*
  • Male
  • Neck / diagnostic imaging*
  • Phantoms, Imaging
  • Signal-To-Noise Ratio*