American Journal of Neuroradiology
American Journal of Neuroradiology

American Journal of Neuroradiology

  • Subscribe
  • Alerts
  • Log in
  • Log out

Advanced Search

  • Home
  • Content
    • Current Issue
    • Publication Preview--Ahead of Print
    • Past Issue Archive
    • Case of the Week Archive
    • Classic Case Archive
    • Case of the Month Archive
    • COVID-19 Content and Resources
  • About Us
    • About AJNR
    • Editorial Board
    • American Society of Neuroradiology
  • For Authors
    • Author Policies
    • Manuscript Submission Guidelines
  • Submit a Manuscript
  • Podcasts
    • Podcasts
    • Subscribe on iTunes
  • More
    • Subscribers
    • Permissions
    • Advertisers
    • Alerts
    • Feedback
  • Follow AJNR on Twitter
  • Visit AJNR on Facebook
  • Follow AJNR on Instagram
  • Join AJNR on LinkedIn
  • RSS Feeds
Research ArticleINTERVENTIONAL

Impact of Intra-Arterial Injection Parameters on Arterial, Capillary, and Venous Time-Concentration Curves in a Canine Model

A.S. Ahmed, Y. Deuerling-Zheng, C.M. Strother, K.A. Pulfer, M. Zellerhoff, T. Redel, K. Royalty, D. Consigny, M.J. Lindstrom and D.B. Niemann
American Journal of Neuroradiology August 2009, 30 (7) 1337-1341; DOI: https://doi.org/10.3174/ajnr.A1586
A.S. Ahmed
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Y. Deuerling-Zheng
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
C.M. Strother
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
K.A. Pulfer
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
M. Zellerhoff
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
T. Redel
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
K. Royalty
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
D. Consigny
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
M.J. Lindstrom
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
D.B. Niemann
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Figures & Data
  • Info & Metrics
  • References
  • PDF
Loading

Abstract

BACKGROUND AND PURPOSE: Recent advances in flat panel detector angiographic equipment have provided the opportunity to obtain physiologic and anatomic information from angiographic examinations. To exploit this possibility, one must understand the factors that affect the bolus geometry of an intra-arterial injection of contrast medium. It was our purpose to examine these factors in a canine model.

MATERIALS AND METHODS: Under an institutionally approved protocol conforming to Guide for the Care and Use of Laboratory Animals of the National Institutes of Health, 7 canines were placed under general anesthesia with isoflurane and propofol. Through a 5F catheter placed into the right common carotid artery, a series of biplane angiographic acquisitions was obtained to examine the effects caused by variation in the volume of injection, the rate of injection, the duration of injection, the concentration of contrast medium, and the catheter position on arterial, capillary, and venous opacification. The results of each injection protocol were determined from analysis of a time-contrast concentration curve derived from locations over an artery, in brain parenchyma, and over a vein. The curve was generated from 2D digital subtraction angiography acquisitions by using prototype software. The area under the curve, the amplitude of the curve, and the time to peak (TTP) were analyzed separately for each injection parameter.

RESULTS: Changes in the injection protocols resulted in predictable changes in the time-concentration curves. The injection parameter that contributed most to maximum opacification was the volume of contrast medium injected. When the injection rate was fixed and the volume was varied, there was an increase in opacification (maximal) proportional to the injected volume. The injected volume also had an indirect (secondary) impact on the temporal characteristics of the opacification. The time-concentration curve became wider, and the peak was shifted to the right as the injection duration increased. The impact of injected volume on maximal opacification was significant (P < .0001), regardless of the site of measurement (artery, tissue, and vein); however, the impact on the temporal characteristics of the time-concentration curve reached statistical significance only in measurements made in the artery and the vein (P < .05), but not in the tissue (P > .1). The impact of injected volume on maximal opacification became nonproportional in the tissue and vein when the volume was very large (>12 mL). Increasing the concentration of contrast medium resulted in a nonproportional increase in the height of the time-concentration curves (P < .05). Injection rate had an impact on both maximal opacification and TTP. The impact on TTP occurred only when the injection rate was very slow (1 mL/s). Changes of concentration had a similar impact on the time-concentration curve. Catheter position did not cause significant alterations in the shape of the curves.

CONCLUSIONS: There were predictable effects from modification of injection parameters on the contrast bolus geometry and on time-concentration curves as measured in an artery, brain parenchyma, or a vein. The amplitude, TTP, and area under the time-concentration curve depend mainly and proportionally on the amount of iodine traversing the vasculature per second. Other injection parameters were of less importance in defining bolus geometry. These findings mimic those observed in studies of parameters affecting bolus geometry following an intravenous injection.

The availability of flat panel detector angiographic equipment has made it possible to perform high-spatial-resolution soft-tissue imaging in the angiographic suite. Recent reports describe the utility of these C-arm CT techniques for the detection of central nervous system hemorrhages, for visualization of endoluminal devices and their relationship to the arterial wall and lumen, and for pretreatment planning and post-treatment evaluations.1–3 Because these applications, along with others that will likely emerge (eg, functional imaging), are often used in association with the intra-arterial administration of iodinated contrast medium, it is necessary to understand the influence of injection parameters that may affect arterial, tissue, and venous opacification.

A variety of parameters may be adjusted when one injects contrast medium. Among others, these include the volume of the injection, the rate of the injection, the duration of the injection, the concentration of the contrast medium that is injected, and the site of injection. Most previous work looking at the effects of variations of these parameters on bolus geometry has been done to try to optimize arterial opacification following an intravenous injection of contrast medium so as to improve the image quality of intravenous digital subtraction angiography (DSA), MR angiography, and CT angiography (CTA).4–7 Although there is very likely a close relationship between the effects of these manipulations on an intra-arterial injection and an intravenous injection in the same subject, to our knowledge, this has not been previously studied.

The purpose of our study was to examine the impact on contrast-bolus geometry after an intra-arterial injection (Table 1) of variation in: 1) the volume of an injection, 2) the rate of an injection, 3) the duration of an injection, 4) the concentration of the contrast injected, and 5) the catheter position.

View this table:
  • View inline
  • View popup
Table 1:

Injection protocol

Materials and Methods

Under an institutionally approved protocol, 7 canines were placed under general anesthesia with isoflurane and propofol. A 5F sheath was placed percutaneously into the right common femoral artery, and through this, a 5F catheter was advanced over a guidewire into the right common carotid artery. There was constant monitoring of heart rate, oxygen saturation, and end-tidal carbon dioxide. Sequential biplane DSA acquisitions were obtained in conjunction with a variety of injection protocols (described below) by using a biplane flat panel detector angiographic system (Artis dBA; Siemens, Erlangen, Germany). Nonionic contrast (300 mg/mL) was power-injected in all studies (Accutron HP-D; Medtron, Saarbrücken, Germany). The angiographic acquisitions were loaded onto a workstation running prototype software (Siemens Healthcare, Forchheim, Germany), which allowed creation of a time-concentration curve of a contrast medium bolus at a selected position on the DSA image. Using these, we could measure specific characteristics and display them on a parametric map. The result from each injection protocol was evaluated by analyzing the time-concentration curves at points in the carotid artery, the torcula, and/or the transverse sinuses and the brain parenchyma. To avoid superimposition of other vascular structures, we made the arterial measurement on an anteroposterior projection, whereas the capillary (tissue) and venous measurements were made on a lateral projection. Care was taken so that the points selected for analysis were such that there was no overlap with other vascular structures (Fig 1). Time-to-peak (TTP) and the area under the curve (AUC) were extracted for the artery, vein, and tissue and were compared. Because there was no movement of the animal or the angiographic table during the acquisitions, the points were the same on all images from each injection protocol. This allowed image-to-image comparison of the curves.

Fig 1.
  • Download figure
  • Open in new tab
  • Download powerpoint
Fig 1.

Three views from the 3D reconstruction used to determine time concentration curves for arterial, parenchyma and venous structures (A, B, and C). The corresponding time-concentration curves are shown in D, E, and F.

We systematically examined 5 variables: 1) volume, 2) rate, 3) duration, 4) concentration, and 5) catheter position, by using the injection protocols summarized in Table 1. Volume, rate, and duration are interrelated variables such that maintaining 1 variable constant requires changing the other 2. The effect of changes in contrast medium concentration was evaluated by using 8 mL of contrast injected at 4 mL/s for 2 seconds. Concentrations of 100%, 75%, 50%, and 25% were tested. This same injection protocol was used to evaluate the effect of varying catheter positions (proximal, middle, and distal) in the right common carotid artery.

TTP, peak opacification, and AUC results from each injection protocol were then evaluated by analyzing the time-concentration curves for the artery (common carotid artery at the bifurcation), vein (torcula), and capillary (brain parenchyma) (Fig 1).

All outcome variables listed in Tables 2–5 were analyzed by using 1-way analysis of variance (ANOVA) after transformation to the log scale to achieve constant variance in the residuals. If the P value for the 1-way ANOVA was significant (P < .05), then pair-wise comparisons of the various levels of the predictor variables were performed by using t tests.

View this table:
  • View inline
  • View popup
Table 2:

P values from the assessment of TTP, peak opacification, and AUC differences when comparing injection protocols: variable rate and volume with constant duration (2 seconds)*

Results

Variations in the injection protocols resulted in predictable changes in the time-concentration curves. The injection parameter that contributed the most to maximum opacification was the injected volume of contrast medium. The impact of this variable on maximal opacification was significant (P < .0001), regardless of the site of measurement (artery, tissue, and vein); however, the impact on the TTP reached statistical significance only with arterial and venous measurements (P < .05). The effect of the contrast volume on the maximal opacification could be observed as well when the injection duration was fixed and both the volume and injection rate varied (Table 2). When the injection rate was fixed, the maximal opacification increased proportionally with the injected volume (P < .01), as long as the injected volume was lower than 12 mL (Table 3). The injected volume had an indirect (secondary) impact on the temporal characteristics (TTP) as well. However, this only occurred in the artery and vein when the injection rate was constant (Table 3 and Fig 2).

Fig 2.
  • Download figure
  • Open in new tab
  • Download powerpoint
Fig 2.

Time-concentration curves observed over an artery (A) and vein (B) at 4 different volumes while the rate of injection remained constant.

View this table:
  • View inline
  • View popup
Table 3:

P values from the assessment of TTP, peak opacification, and AUC differences when comparing injection protocols: variable volume and duration with constant rate*

Alterations to the contrast concentration yielded results similar to those of altering the injection volume, with higher concentrations paralleling the relationship of higher injected contrast volumes. This effect was most significant in the arterial and venous TTP and in all 3 measurements for peak opacification when comparing full strength (100%) with dilute (25%) contrast medium (Table 4).

View this table:
  • View inline
  • View popup
Table 4:

P values from the assessment of TTP, peak opacification, and AUC differences when comparing injection protocols: variable concentration*

Injection rate had an impact on both maximal opacification and TTP. The greatest impact on TTP occurred when comparing the slowest (1 mL/s) injection rate with the other rates (Table 5). On the other hand, when the injected volume was constant, the TTP was the most stable (Table 5).

View this table:
  • View inline
  • View popup
Table 5:

P values from the assessment of TTP, peak opacification, and AUC differences when comparing injection protocols: variable rate and duration with constant volume*

Discussion

Studies of injection protocols on bolus geometry have previously analyzed the effects of injection parameters on arterial opacification following an intravenous injection.8,9 Our study assessed the impact of injection parameters on an intra-arterial bolus of contrast medium, looking not only at the effects on arterial opacification but also at opacification of parenchyma and venous structures.

The bolus geometry measured in the arterial capillary and venous phases represents the effects of the injection parameters modulated by the subject's physiology (eg, patient function). Because of this, measurements made near the site of injection (artery), where the subject's physiology has not yet exerted a full effect on the bolus, result in intensity time curves that are more striking than those in the capillaries (tissue) or veins.

Interpretation of independent variables such as contrast concentration and catheter position could be assessed individually. For the inter-related variables of volume, rate, and duration, analysis of all 3 led to an understanding of the effect of each variable. Many of the results were predictable. The more contrast that is given in a shorter time period, the greater and earlier is the peak opacification. This is more easily seen in the artery, with a greater change in the parameters being required to observe a difference in the capillaries and vein. Similar curves can be obtained by changing either the volume of the contrast injected or by altering the contrast concentration so that the total number of iodine molecules is similar.

Our results parallel previously published results of an intravenous-injection-parameter effect on intra-arterial opacification: The volume of the contrast injected causes the greatest effect on arterial, capillary, and venous opacification.1 Cademartiri et al5 summarized the effects of varying injection parameters in a review of intravenous injections for CTA. In our study using intra-arterial injections, we observed similar effects, especially when measurements were made in the artery.

As functional imaging methods are developed to measure perfusion parameters by using C-arm CT, it will be imperative to have an understanding of the effect of altering injection parameters on the bolus geometry. Due to the poor temporal resolution of C-arm CT, a bolus-tracking methodology, as used in perfusion CT, is currently not feasible. Instead, a relatively constant (steady) level of contrast medium must be maintained in the tissue for perfusion values to be valid.

Our observations, thus, have clinical significance in that they confirm the close relationships between the effects of manipulations of injection parameters on an intra-arterial injection and an intravenous injection in the same subject. As developments in perfusion measurements using C-arm CT evolve, it seems likely that serial measurements of perfusion in local areas of a particular tissue or organs will be facilitated through the use of an intra-arterial injection. This would further increase the value of understanding the interactions that we describe. Choosing a set of optimal parameters will depend on the goal of a particular imaging acquisition and on the subject's individual physiology.4

Conclusions

Contrast volume was the parameter that had the strongest impact on the time-concentration curve. It had more impact on the maximal opacification than on the temporal characteristics. The impact on maximal opacification diminished when the volume was very large (>12 mL). The injection rate impacted both maximal opacification and temporal characteristics. When the volume was fixed, the impact on TTP occurred only when the injection rate was very slow (1 mL). With regard to the variation of the time-concentration curve, the injection protocol with constant volume seemed to be the most stable.

References

  1. ↵
    Heran HS, Song JK, Namba K, et al.The utility of DynaCT in neuroendovascular procedures. AJNR Am J Neuroradiol 2006;27:330–32
    Abstract/FREE Full Text
  2. Benndorf G, Strother CM, Claus B, et al. Angiographic CT in cerebrovascular stenting. AJNR Am J Neuroradiol 2005;26:1813–18
    Abstract/FREE Full Text
  3. ↵
    Wallace MJ, Kuo MD, Glaiberman C, et al., Three-dimensional C-arm cone-beam CT: applications in the interventional suite. J Vasc Interv Radiol 2008;19:799–813
    CrossRefPubMed
  4. ↵
    Burbank FH, Brody WR, Bradley BR. Effect of volume and rate of contrast medium injection on intravenous digital subtraction angiographic contrast medium curves. J Am Coll Cardiol 1984;4:308–15
    PubMed
  5. ↵
    Cademartiri F, van der Lugt A, Luccichenti G, et al. Parameters affecting bolus geometry in CTA: a review. J Comput Assist Tomogr 2002;26:598–607
    CrossRefPubMed
  6. Bae KT. Peak contrast enhancement in CT and MR angiography: When does it occur and why? Pharmacokinetic study in a porcine model. Radiology 2003;227:809–16
    PubMed
  7. ↵
    Vrachliotis TG, Kostaki GB, Ahmad H, et al. Atypical chest pain: coronary, aortic and pulmonary vasculature enhancement at biphasic single-injection 64-section CT angiography. Radiology 2007;243:368–76
    CrossRefPubMed
  8. ↵
    Claussen CD, Banzer D, Pfretzschner C, et al. Bolus geometry and dynamics after intravenous contrast medium injection. Radiology 1984;153:365
    PubMed
  9. ↵
    Reiser UJ. Study of bolus geometry after intravenous contrast medium injection: dynamic and quantitative measurements (Chronogram) using an x-ray CT device. J Comput Assist Tomogr 1984 :8:251–62
    PubMed
  • Received November 5, 2008.
  • Accepted after revision February 16, 2009.
  • American Society of Neuroradiology
View Abstract
PreviousNext
Back to top

In this issue

American Journal of Neuroradiology: 30 (7)
American Journal of Neuroradiology
Vol. 30, Issue 7
August 2009
  • Table of Contents
  • Index by author
Advertisement
Print
Download PDF
Email Article

Thank you for your interest in spreading the word on American Journal of Neuroradiology.

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Impact of Intra-Arterial Injection Parameters on Arterial, Capillary, and Venous Time-Concentration Curves in a Canine Model
(Your Name) has sent you a message from American Journal of Neuroradiology
(Your Name) thought you would like to see the American Journal of Neuroradiology web site.
CAPTCHA
This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.
Citation Tools
Impact of Intra-Arterial Injection Parameters on Arterial, Capillary, and Venous Time-Concentration Curves in a Canine Model
A.S. Ahmed, Y. Deuerling-Zheng, C.M. Strother, K.A. Pulfer, M. Zellerhoff, T. Redel, K. Royalty, D. Consigny, M.J. Lindstrom, D.B. Niemann
American Journal of Neuroradiology Aug 2009, 30 (7) 1337-1341; DOI: 10.3174/ajnr.A1586

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
Share
Impact of Intra-Arterial Injection Parameters on Arterial, Capillary, and Venous Time-Concentration Curves in a Canine Model
A.S. Ahmed, Y. Deuerling-Zheng, C.M. Strother, K.A. Pulfer, M. Zellerhoff, T. Redel, K. Royalty, D. Consigny, M.J. Lindstrom, D.B. Niemann
American Journal of Neuroradiology Aug 2009, 30 (7) 1337-1341; DOI: 10.3174/ajnr.A1586
del.icio.us logo Digg logo Reddit logo Twitter logo CiteULike logo Facebook logo Google logo Mendeley logo
  • Tweet Widget
  • Facebook Like
  • Google Plus One
Purchase

Jump to section

  • Article
    • Abstract
    • Materials and Methods
    • Results
    • Discussion
    • Conclusions
    • References
  • Figures & Data
  • Info & Metrics
  • References
  • PDF

Related Articles

  • No related articles found.
  • PubMed
  • Google Scholar

Cited By...

  • Flat Detector CT in the Evaluation of Brain Parenchyma, Intracranial Vasculature, and Cerebral Blood Volume: A Pilot Study in Patients with Acute Symptoms of Cerebral Ischemia
  • Parametric Color Coding of Digital Subtraction Angiography
  • Crossref (20)
  • Google Scholar

This article has been cited by the following articles in journals that are participating in Crossref Cited-by Linking.

  • Parametric Color Coding of Digital Subtraction Angiography
    C.M. Strother, F. Bender, Y. Deuerling-Zheng, K. Royalty, K.A. Pulfer, J. Baumgart, M. Zellerhoff, B. Aagaard-Kienitz, D.B. Niemann, M.L. Lindstrom
    American Journal of Neuroradiology 2010 31 5
  • Flat Detector CT in the Evaluation of Brain Parenchyma, Intracranial Vasculature, and Cerebral Blood Volume: A Pilot Study in Patients with Acute Symptoms of Cerebral Ischemia
    T. Struffert, Y. Deuerling-Zheng, S. Kloska, T. Engelhorn, C.M. Strother, W.A. Kalender, M. Köhrmann, S. Schwab, A. Doerfler
    American Journal of Neuroradiology 2010 31 8
  • Parametric Color Coding of Digital Subtraction Angiography in the Evaluation of Carotid Cavernous Fistulas
    P. Gölitz, T. Struffert, H. Lücking, J. Rösch, F. Knossalla, O. Ganslandt, Y. Deuerling-Zheng, A. Doerfler
    Clinical Neuroradiology 2013 23 2
  • Monitoring of Balloon Test Occlusion of the Internal Carotid Artery by Parametric Color Coding and Perfusion Imaging Within the Angio Suite: First Results
    T. Struffert, Y. Deuerling-Zheng, T. Engelhorn, S. Kloska, P. Gölitz, A. Bozzato, M. Kapsreiter, C. M. Strother, A. Doerfler
    Clinical Neuroradiology 2013 23 4
  • Measurement of quantifiable parameters by time-density curves in the elastase-induced aneurysm model: first results in the comparison of a flow diverter and a conventional aneurysm stent
    Tobias Struffert, Sabine Ott, Markus Kowarschik, Frederik Bender, Edyta Adamek, Tobias Engelhorn, Philipp Gölitz, Stefan Lang, Charles M. Strother, Arnd Doerfler
    European Radiology 2013 23 2
  • Quantitative assessment of angiographic perfusion reduction using color-coded digital subtraction angiography during transarterial chemoembolization
    Ji Wang, Jie-Jun Cheng, Kai-Yi Huang, Zhi-Guo Zhuang, Xue-Bin Zhang, Jia-Chang Chi, Xiao-Lan Hua, Jian-Rong Xu
    Abdominal Radiology 2016 41 3
  • Where Contrast Agent Concentration Really Matters – A Comparison of CT and MRI
    Sabine Heiland, Gunter Erb, Simon Ziegler, Martin Krix
    Investigative Radiology 2010 45 9
  • New grading of moyamoya disease using color-coded parametric quantitative digital subtraction angiography
    Sheng-Che Hung, Muh-Lii Liang, Chun-Fu Lin, Chung-Jung Lin, Wan-Yuo Guo, Feng-Chi Chang, Tai-Tong Wong, Cheng-Yen Chang
    Journal of the Chinese Medical Association 2014 77 8
  • Objective Assessment of Transcatheter Arterial Chemoembolization Angiographic Endpoints: Preliminary Study of Quantitative Digital Subtraction Angiography
    Xue-bin Zhang, Zhi-guo Zhuang, Hui Ye, Janina Beilner, Markus Kowarschik, Jie-jun Chen, Jia-chang Chi, Jian-rong Xu
    Journal of Vascular and Interventional Radiology 2013 24 5
  • A model based algorithm for perfusion estimation in interventional C‐arm CT systems
    Martin Wagner, Yu Deuerling‐Zheng, Markus Möhlenbruch, Martin Bendszus, Jan Boese, Sabine Heiland
    Medical Physics 2013 40 3

More in this TOC Section

  • Partial (SAVE) versus Complete (Solumbra) Stent Retriever Retraction Technique for Mechanical Thrombectomy: A Randomized In Vitro Study
  • Contrast Injection from an Intermediate Catheter Placed in an Intradural Artery is Associated with Contrast-Induced Encephalopathy following Neurointervention
  • CT Perfusion Does Not Modify the Effect of Reperfusion in Patients with Acute Ischemic Stroke Undergoing Endovascular Treatment in the ESCAPE-NA1 Trial
Show more INTERVENTIONAL

Similar Articles

Advertisement

News and Updates

  • Lucien Levy Best Research Article Award
  • Thanks to our 2022 Distinguished Reviewers

Resources

  • Evidence-Based Medicine Level Guide
  • AJNR Podcast Archive
  • Librarian Resources
  • Terms and Conditions

Opportunities

  • Get Peer Review Credit from Publons

American Society of Neuroradiology

  • Neurographics
  • ASNR Annual Meeting
  • Fellowship Portal

© 2023 by the American Society of Neuroradiology | Print ISSN: 0195-6108 Online ISSN: 1936-959X

Powered by HighWire