Abstract
SUMMARY: To determine how much the radiation dose can be reduced in multidetector row CTA using a QDS, we performed CTA at various exposure settings using a vascular phantom simulating various aneurysms with superimposed bone skull structures, and postprocessed the image data with QDS. Our results demonstrated that the radiation dose of CTA can be reduced by at least 25% and the image quality for visualizing aneurysms can be preserved by applying the QDS.
Abbreviations
- CNR
- contrast-to-noise ratio
- CTA
- CT angiography
- CTDIw
- weighted CT dose index
- MTF
- modulation transfer function
- Q04
- mild QDS
- Q06
- heavy sharpening QDS
- Q08
- heavy smoothing QDS
- QDS
- quantum denoising system
- ROI
- region of interest
- W
- weight of blending the smoothed image and the sharpened image