Abstract
BACKGROUND AND PURPOSE: Blood flow management in the carotid artery during mechanical thrombectomy is crucial for safety and effectiveness. There is an ongoing discussion about whether balloon-guide catheters or large-bore sheaths are needed for effective flow management. We compared general flow characteristics of proximal aspiration through a large-bore sheath and a balloon-guide catheter in a porcine in vivo model.
MATERIALS AND METHODS: We investigated blood flow in a porcine common carotid artery with and without aspiration (VacLok syringe and Penumbra pump, Pump MAX) through an 8F-long sheath and an 8F balloon-guide catheter. Blood hemodynamics were assessed via continuous duplex sonography.
RESULTS: Average vessel diameter and baseline blood flow were 4.4 ± 0.2 mm and 244 ± 20 mL/min, respectively. For the 8F sheath, pump aspiration resulted in a significant flow reduction (225 ± 25 mL/min, P < .001), but with a persisting antegrade stream. Manual aspiration resulted in collapse of the vessel in 2 of 7 measurements and oscillatory flow with antegrade systolic and retrograde diastolic components in the remaining 5 measurements. Net flow was antegrade (52 ± 44 mL/min) in 3 and retrograde (−95 ± 52 mL/min) in the remaining 2 measurements. For balloon-guide catheters, balloon inflation always resulted in flow arrest. Additional pump or manual aspiration resulted in significant flow reversal of −1100 ± 230 and −468 ± 46 mL/min, respectively (both, P < .001).
CONCLUSIONS: Only balloon-guide catheters allow reliable blood flow arrest and flow reversal in combination with aspiration via syringes or high-flow pump systems. Aspiration through an 8F sheath results in either collapse of the vessel or oscillatory flow, which can result in a net antegrade or retrograde stream.
ABBREVIATIONS:
- BGC
- balloon-guide catheter
- CCA
- common carotid artery
Endovascular mechanical thrombectomy has been established as the standard treatment option for acute ischemic stroke caused by large-vessel occlusion.1 Even though the basic principle of mechanical thrombectomy is established, actual procedures differ in many ways, ranging from the choice of stent retrievers to the choice of access catheters.2⇓–4 Recent clinical studies imply better procedural and clinical outcomes for patients who are treated with balloon-guide catheters (BGCs) compared with other carotid-access catheters such as large-bore sheaths.5,6 Zaidat et al5 found, in their analysis of the Systematic Evaluation of Patients Treated With Stroke Devices for Acute Ischemic Stroke (STRATIS) Registry data, that BGCs are associated with higher rates of good clinical outcome (mRS ≤ 2), successful revascularization, a trend toward higher rates of complete first-pass revascularization, and a lower number of overall passes.7 Because these data are not derived from randomized prospective trials, it is possible that the superiority of BGCs simply reflects their being used by more experienced interventionalists. Nonetheless, it is reasonable that blood flow management in the carotid artery is crucial for safety and effectiveness during mechanical thrombectomy.8 It has been shown that thrombectomy maneuvers produce thousands of small clot fragments that may occlude small arterioles and capillaries.4 The result can be small infarctions that can have a relevant clinical impact despite being invisible on MR imaging.9,10 Hence, flow arrest is important regardless of the specific thrombectomy technique (eg, classic stent-retriever thrombectomy, A Direct Aspiration First Pass Technique [ADAPT], Solumbra, or other techniques). Antegrade blood flow should be reversed during mechanical thrombectomy to prevent embolization of clot fragments.8 This is usually attempted via blood aspiration through an access catheter in the internal carotid artery. There is a variety of access catheters, ranging from small (5F–7F) guiding catheters to large-bore sheath catheters (8F) and BGCs, but there is no consensus about which access catheter is most suitable.4,8,11⇓–13
Mechanical thrombectomy is becoming more common, and many interventionalists must decide which technique to choose. Because empiric data are lacking and large randomized prospective studies are needed to resolve this issue from a clinical point of view, we addressed this question with a simple technical approach. To reflect basic principles of flow management in real-life settings, we quantified and compared flow characteristics and aspiration volumes of proximal aspiration through a large-bore sheath and a BGC with both manual and pump aspiration. Because it is practically impossible to quantify blood flow during mechanical thrombectomy in the ICA, let alone in the cerebral arteries, we decided to investigate flow characteristics in an in vivo porcine model.
Materials and Methods
All experiments were performed on 4 female Landrace swine (weight, 50–60 kg) with peri- and intrainterventional management as reported previously.14 The experiments were performed in accordance with the German legislation governing animal studies following the “Guide for the Care and Use of Laboratory Animals” (https://grants.nih.gov/grants/olaw/Guide-for-the-Care-and-use-of-laboratory-animals.pdf) and the “Directive 2010/63/EU on the Protection of Animals Used for Scientific Purposes” (https://www.researchgate.net/publication/233428185_DIRECTIVE_201063EU_on_the_protection_of_animals_used_for_scientific_purposes). Official permission was granted from the governmental animal care and use office, Landesamt für Natur, Umwelt und Verbraucherschutz Nordrhein-Westfalen, Recklinghausen, Germany.
We investigated flow in a porcine common carotid artery (CCA) with and without aspiration through a large-bore sheath and a BGC. We chose the CCA because its diameter is comparable with that of the human ICA.15 We adapted flow in the CCA, which has a physiologic blood flow that is twice as high the human ICA blood flow, by injecting blood clots into the subsequent arteries. This procedure was repeated until a blood flow between 200 and 280 mL/min was achieved in the target vessel. We assessed blood flow via duplex sonography (LOGIQ S8; GE Healthcare, Milwaukee, Wisconsin) by measuring vessel diameter and flow speed using the built-in software and GraphClick software (Arizona Software, Neuchâtel, Switzerland).
Instead of testing all possible access catheters ranging from 5F guide catheters to large-bore sheaths, we chose 1 large-bore sheath and 1 large-bore BGC to investigate flow characteristics in 2 setups that supposedly allow maximum flow control. We used an 8F-long sheath, Shuttle Select (Cook, Bloomington, Indiana; outer diameter, 3 mm; inner diameter, 2.87 mm) and an 8F FlowGate II BGC (Concentric Medical, Mountain View, California; outer diameter, 2.7 mm; inner diameter, 2.1 mm), which are currently the largest of the most commonly used proximal access catheters. Sheaths and BGCs were introduced through a femoral access and placed 5–7 cm behind the CCA origin. Blood flow was assessed 7–10 cm distal to the sheath/BGC tips (Fig 1). We did not measure immediately behind the catheter tips to avoid measuring blood flow in areas of turbulent flow. After quantifying the baseline blood flow in the CCA, we quantified blood flow with and without aspiration through the sheath and the inflated BGC. Aspiration was performed manually with a 60-mL VacLok syringe (Merit Medical Systems, South Jordan, Utah) and with a Penumbra pump, Pump MAX (Penumbra, Alameda, California), which is the most common aspiration pump, with the recommended vacuum pressure of −25.5 Hg. All experiments were performed at least 7 times, and all measurements were included in our final analysis. Because our experiments resulted in considerable blood loss, not all experiments were repeated in all swine. However, every setup was performed in at least 2 swine to increase the variance. To restrict aspirated blood volume, we performed only 3 aspiration experiments with an inflated BGC as a proof of principle. Blood flow was assessed regularly between experiments to anticipate unnoticed shifts of baseline values.
Statistical Analysis
Student t tests were used for comparison of flow volumes after testing our data for normal distribution with a Shapiro-Wilk test. P values of an α level of ≤.05 were significant. All statistical analyses were performed with SPSS 23 software (IBM, Armonk, New York).
Results
The average diameter of the CCA and baseline blood flow were 4.4 ± 0.2 mm and 244 ± 20 mL/min, respectively (Fig 1). The 8F sheath resulted in an average reduction of vessel diameter and cross-sectional area of 61% and 37%, respectively, and the sheath did not significantly reduce blood flow (236 ± 25 mL/min, P = .19). Pump aspiration resulted in a significant flow reduction with a net flow of 225 ± 25 mL/min (P < .001) (Fig 2). However, flow always remained antegrade, and there was no flow arrest or reversal. Manual aspiration through the 8F sheath resulted in collapse of the vessel in 2 of 7 measurements. In the remaining 5 measurements, there was an oscillatory flow with antegrade flow in systole and retrograde flow in diastole (Fig 3). Net flow was antegrade (52 ± 44 mL/min) in 3 of 5 measurements and retrograde (−95 ± 52 mL/min) in the remaining 2 measurements.
The noninflated 8F BGC in the CCA led to an average reduction of vessel diameter and cross-sectional area of 61% and 37%, respectively, and slightly reduced blood flow (from 244 ± 20 to 228 ± 22 mL/min, P = .02). Balloon inflation always resulted in arrest of antegrade flow (Fig 4). Additional pump aspiration and manual aspiration through an inflated BGC resulted in a constant and significant flow reversal with a flow of −1100 ± 230 and −468 ± 46 mL/min, respectively (both, P < .001) (Fig 4). Net flow in the carotid artery was significantly lower when blood was aspirated through an inflated BGC compared with the 8F sheath, regardless of pump or syringe aspiration (both, P < .001).
Discussion
Ever since mechanical stroke treatment has been established as a standard treatment technique, the focus of neurointerventional stroke research has shifted to finding the optimal treatment techniques.1,2,4,16⇓–18 Proximal aspiration in the ICA to achieve flow reversal has been established as a standard technique for mechanical thrombectomy. In an average patient and without application of a balloon-guide catheter, which restricts antegrade flow, the amount of aspirated blood in the ICA should surpass its normal flow of 240 mL/min to achieve reliable flow reversal.19 There is an ongoing discussion of whether BGCs or large-bore sheaths are needed for effective flow management. Large-bore sheaths allow aspiration of high-flow volumes up to approximately 1000 mL/min.20 However, the handling of these sheaths is rather complicated because their stiffness necessitates additional guiding catheters for placement. BGCs allow reliable occlusion of the respective artery and consecutive reduction of antegrade flow. However, rather difficult handling and their comparably small inner diameters, which restrict the choice of catheters/devices that can be introduced through the BGC, prevent many interventionalists from using BGCs.
Our results show that the choice of carotid access catheter has in important impact on blood flow in the target vessel. We have found that only BGCs allow reliable flow arrest in our in vivo porcine model. In a patient, this does not mean that balloon inflation results in arrest of cerebral perfusion because collaterals may maintain cerebral blood flow distal to the occlusion site. As long as the clot is located above the circle of Willis, flow from the posterior and anterior communicating arteries and the ophthalmic artery can sustain the full physiologic flow volume of the ICA and may cause embolization of clot fragments.21 Consequently, aspiration of blood through an inflated BGC is necessary for reliable retrograde flow and control of peripheral embolization. Aspiration through the BGC with the syringe and with the Penumbra pump resulted in constant retrograde flow. However, only the syringe achieved high flow that surpassed the required antegrade flow rate of 240 mL/min for reliably reverting blood flow at the level of the terminal segment of the internal carotid artery. Therefore, we discourage using the Penumbra pump if flow reversal during mechanical thrombectomy is intended.
Surprisingly, aspiration with large-bore sheaths did not necessarily result in flow arrest or reversal, despite high-aspiration flow volumes. Aspiration through an 8F sheath with a syringe or a Penumbra pump results in flow volumes of ≈800 and ≈240 mL/min, respectively.20 By simply adding antegrade and retrograde flow, we would have expected a net backward flow of >500 mL/min with the syringe and flow arrest with the Penumbra pump. However, aspiration through an 8F sheath did not result in reliable flow reversal. At best, there was oscillatory flow with small net backward flow volumes, which would not suffice to reverse blood flow above the circle of Willis. This behavior is likely because aspiration results in an additional pressure gradient that mobilizes additional flow from the aortic arch into the CCA along the catheter while antegrade flow in the ICA is maintained (Fig 2). In the worst-case scenario, aspiration resulted in vessel collapse, which coincides with our clinical experience: During our interventions, we have found that when too much suction is applied, blood cannot be aspirated until the suction is decreased either because the catheter tip has engaged the vessel wall or—as we surmise—the vessel collapsed. Vessel collapse in perfused vessels is mainly due to the Bernoulli effect, which causes the vessel walls to converge during aspiration of high flow volumes. This issue is relevant in clinical practice because it is almost impossible to predict vessel collapse: Many factors such as flow volume, position of the catheter, occlusion site, and the diameter of the vessel and its rigidity have an impact on the applied forces that cause vessel collapse. Given that aspiration with large-bore sheaths did not reliably result in flow arrest or reversal, we discourage using large-bore catheters or sheaths for flow arrest in the ICA, regardless of syringe or pump aspiration. Furthermore, we specifically discourage using smaller guiding catheters (5F–7F), which are very common because of their easy handling and low cost but face the same issue of insufficient aspiration like the larger sheath tested in our experiment. Notably, even under ideal conditions (perfect vacuum and no external resistors), aspiration flow through a 6F catheter with a stent retriever in its lumen does not reach 240 mL/min.17
Limitations
A major limitation of our study is that it does not allow assessing whether cerebral blood flow above the circle of Willis is reversed. Such an experiment would need transcranial Doppler measurements in patients during thrombectomy, which is impossible due to the artifacts caused by the microcatheter and stent retriever in place. As an approximation, we estimated the following: Physiologic blood flow in an internal carotid artery is approximately 240 mL/min. It has been shown that aspiration flow through an 8F BGC with pump and syringe aspiration can surpass 350 mL/min (pump) and 500 mL/min (syringe), which again should be sufficient to allow flow arrest despite collateral flow. Nonetheless, there remains the possibility that collateral flow through the circle of Willis may be enough to maintain antegrade cerebral blood flow despite high-aspiration flow volumes.20 Also, due to the nature of our study, we could not investigate whether flow reversal has an effect on clinical outcome, but our results may serve as a foundation for future research that specifically addresses this question. Also, even though the porcine cardiovascular system serves as an excellent model, it may be arguable whether our quantitative results can be translated to patients without restrictions. Nonetheless, because it is impossible to conduct such an experiment in patients the porcine model is the best approximation to real life.22 Last, flow volume measurements based on duplex sonography may not always reflect actual flow volumes and are a minor limitation of our study.23 However, even if flow volumes are over- or underestimated, our results remain valid on a qualitative level because the direction of flow is unaffected by the measuring method.
Conclusions
Only BGCs allow reliable flow reversal and should be used with syringes or high-flow pump systems. Aspiration through an 8F sheath in the ICA results in either collapse of the vessel or oscillatory flow with antegrade flow in systole and retrograde flow in diastole, which can result in a net antegrade or retrograde flow.
Footnotes
Disclosures: Martin Wiesmann—UNRELATED: Consultancy: Stryker Neurovascular; Payment for Lectures Including Service on Speakers Bureaus: Bracco, Medtronic, Siemens, Stryker Neurovascular; Payment for Development of Educational Presentations: Abbott, ab medica, Acandis, Bayer HealthCare, Bracco, B. Braun, Codman Neurovascular, Kaneka Pharma Europe, Medtronic, Dahlhausen Medizintechnik, MicroVention, Penumbra, phenox, Philips Healthcare, Route 92, Siemens, Silk Road Medical, St. Jude, Stryker Neurovascular.* *Money paid to the institution.
References
- Received April 30, 2018.
- Accepted after revision August 13, 2018.
- © 2018 by American Journal of Neuroradiology