PT - JOURNAL ARTICLE AU - Jablonska, Magda AU - Li, Jiahui AU - Tiberi, Riccardo AU - Canals, Pere AU - Ortega, Santiago AU - Tomasello, Alejandro AU - Ribo, Marc TI - Partial (SAVE) versus Complete (Solumbra) Stent Retriever Retraction Technique for Mechanical Thrombectomy: A Randomized In Vitro Study AID - 10.3174/ajnr.A7996 DP - 2023 Oct 01 TA - American Journal of Neuroradiology PG - 1165--1170 VI - 44 IP - 10 4099 - http://www.ajnr.org/content/44/10/1165.short 4100 - http://www.ajnr.org/content/44/10/1165.full SO - Am. J. Neuroradiol.2023 Oct 01; 44 AB - BACKGROUND AND PURPOSE: Mechanical thrombectomy has become a first-line treatment for acute ischemic stroke. Several techniques combining stent retrievers and distal aspiration catheters have been described. We aimed to characterize the efficacy of 2 commonly used techniques according to clot characteristics.MATERIALS AND METHODS: Soft (mean stiffness = 95.77 [SD, 5.80] kPa) or stiff (mean stiffness = 205.63 [SD, 6.70] kPa) clots (3 × 10 mm and 2 × 10 mm, respectively) were embolized to the distal M1 segment of the MCA in an in vitro model. The technique was randomly allocated (1:1): stent retriever assisted vacuum-locked extraction (SAVE) versus complete retraction (Solumbra). The primary end point was the percentage of first-pass recanalization. Secondary end points were periprocedural distal embolization measures.RESULTS: A total of 130 mechanical thrombectomies were performed (50 for soft clots and 15 for stiff clots per arm). Overall, the rate of first-pass recanalizaton was 35% with Solumbra and 15% with SAVE (P < .01). For stiff clots, the first-pass recanalizaton was equal for both methods (27%; P = 1.00). With soft clots, the first-pass recanalizaton was higher with Solumbra (38%) than with SAVE (12%; P < .01). When we used soft clots, the maximum embolus size (mean, 1.19 [SD, 0.9] mm versus 2.16 [SD, 1.48] mm; P < .01) and total area of emboli (mean, 1.82 [SD, 2.73] versus 3.34 [SD, 3.2]; P = .01) were also lower with Solumbra than with SAVE.CONCLUSIONS: Clot characteristics may influence the efficacy of the thrombectomy technique. In occlusions caused by soft clots, complete retrieval into the distal aspiration catheters achieved higher rates of first-pass recanalizaton and lower embolization.DACdistal aspiration catheter%FPRpercentage of first-pass recanalizationMTmechanical thrombectomySAVEstent retriever assisted vacuum-locked extractionSolumbratotal retractionSRstent retriever