RT Journal Article SR Electronic T1 3D High-Spatial-Resolution Cerebral MR Venography at 3T: A Contrast-Dose-Reduction Study JF American Journal of Neuroradiology JO Am. J. Neuroradiol. FD American Society of Neuroradiology SP 349 OP 355 DO 10.3174/ajnr.A1319 VO 30 IS 2 A1 A. Tomasian A1 N. Salamon A1 M.S. Krishnam A1 J.P. Finn A1 J.P. Villablanca YR 2009 UL http://www.ajnr.org/content/30/2/349.abstract AB BACKGROUND AND PURPOSE: The effect of various contrast-dose regimens for cerebral MR venography (MRV) has not been previously evaluated at 3T, to our knowledge. Our purpose was to evaluate and compare the diagnostic image quality resulting from half-versus-full-dose contrast regimens for high-spatial-resolution 3D cerebral MRV at 3T.MATERIALS AND METHODS: Forty consecutive patients with known or suggested cerebrovascular disease underwent 3D high-spatial-resolution (0.7 × 0.6 × 0.9 mm3) cerebral contrast-enhanced MRV (CE-MRV) at 3T, by using an identical acquisition protocol. Patients were assigned to 1 of 2 groups: 1) full-dose (∼0.1 mmol/kg), and 2) half-dose (∼0.05 mmol/kg). Two readers evaluated the resulting images for overall image quality, venous structure definition, and arterial contamination. Signal intensity–to-noise-ratio (SNR) and contrast-to-noise-ratio (CNR) were evaluated in 8 consistent sites. Statistical analysis was performed by using Mann-Whitney U, Wilcoxon signed rank, and t tests and a κ coefficient.RESULTS: Both readers scored venous-structure definition as excellent or sufficient for diagnosis in approximately 90% of segments for the full-dose group (κ = 0.87) and in approximately 80% of segments for the half-dose group (κ = 0.85). Delineation grades were significantly lower for small venous segments, including the middle cerebral, septal, superior cerebellar, inferior vermian, posterior tonsillar, and thalamostriate veins in the half-dose group (P < .01). No significant difference existed for arterial contamination grades between the 2 groups (P > .05). SNR and CNR values were lower in the half-dose group (P < .01).CONCLUSIONS: At 3T, high-spatial-resolution cerebral MRV can be performed with contrast doses as low as 7.5 mL, without compromising image quality as compared with full-dose protocols, except in the smallest veins, and without compromise of acquisition speed or spatial resolution.