PT - JOURNAL ARTICLE AU - W.P. Daugherty AU - J.B. White AU - H.J. Cloft AU - D.F. Kallmes TI - Rescue Retrieval of AngioGuard Embolic Capture System after Failure of Capture Sheath Retrieval in Carotid Angioplasty and Stenting AID - 10.3174/ajnr.A1118 DP - 2008 Sep 01 TA - American Journal of Neuroradiology PG - 1594--1595 VI - 29 IP - 8 4099 - http://www.ajnr.org/content/29/8/1594.short 4100 - http://www.ajnr.org/content/29/8/1594.full SO - Am. J. Neuroradiol.2008 Sep 01; 29 AB - SUMMARY: This report describes the cases of 2 patients with failed retrieval of an AngioGuard distal protection device after carotid stent placement with rescue retrieval by use of a vertebral catheter. An AngioGuard was deployed before carotid angioplasty and stent placement. Attempted removal of the device with the provided recapture sheath was unsuccessful. A 5F 120-cm vertebral catheter was used to recapture and remove the AngioGuard. This technique is a simple and readily available solution for the retrieval of failed removal of a protection device.