RT Journal Article SR Electronic T1 Rescue Retrieval of AngioGuard Embolic Capture System after Failure of Capture Sheath Retrieval in Carotid Angioplasty and Stenting JF American Journal of Neuroradiology JO Am. J. Neuroradiol. FD American Society of Neuroradiology SP 1594 OP 1595 DO 10.3174/ajnr.A1118 VO 29 IS 8 A1 W.P. Daugherty A1 J.B. White A1 H.J. Cloft A1 D.F. Kallmes YR 2008 UL http://www.ajnr.org/content/29/8/1594.abstract AB SUMMARY: This report describes the cases of 2 patients with failed retrieval of an AngioGuard distal protection device after carotid stent placement with rescue retrieval by use of a vertebral catheter. An AngioGuard was deployed before carotid angioplasty and stent placement. Attempted removal of the device with the provided recapture sheath was unsuccessful. A 5F 120-cm vertebral catheter was used to recapture and remove the AngioGuard. This technique is a simple and readily available solution for the retrieval of failed removal of a protection device.