ADC measurement reproducibility for the different subgroups
Variable | Mean (×10−6 mm2/s) | Within-Subject SD | 95% Reference Rangea | CV (%)b | Reliability Coefficientc |
---|---|---|---|---|---|
Analysis/re-analysis: | |||||
Control mean (n = 10) | 928 | 23 | ±45 | 2.5 | 0.99 |
Diseased optic nerves (n = 16) | 1324 | 97 | ±189 | 7.3 | 0.94 |
Healthy contralateral optic nerves (n = 14) | 990 | 31 | ±61 | 3.2 | 0.96 |
Image/reimage: | |||||
Control mean (n = 7) | 946 | 95 | ±186 | 10.0 | 0.58 |
a 1.96 × within-subject SD; 95% of measurements are expected to lie within this departure from the true value.
b CV, coefficient of variation.
c The proportion of total variance due to between-subject variation. Under assumptions that are plausible here, 1 minus this value is the proportion of variation due to measurement error.