Table 3:

Diagnostic performance according of each observer regarding the detection of the simulated aneurysmal blebs

Contrast Material Concentration & Dose ModeObserverMean Az
1234567
300 mgI/mL
    Standard-dose mode of FPD system0.930.910.890.920.910.920.940.92
    Low-dose mode of FPD system0.890.890.910.920.920.940.950.92
    Conventional-dose mode of I.I.-TV system0.920.910.860.980.870.950.900.91
150 mgI/mL
    Standard-dose mode of FPD system0.810.790.830.830.830.800.820.82
    Low-dose mode of FPD system0.830.790.850.820.830.850.840.83
    Conventional-dose mode of I.I.-TV system0.800.780.850.810.820.840.820.82
100 mgI/mL
    Standard-dose mode of FPD system0.840.880.870.800.930.870.800.86*
    Low-dose mode of FPD system0.840.880.860.790.930.870.790.85*
    Conventional-dose mode of I.I.-TV system0.670.550.540.570.550.540.520.56
  • Note:—Az indicates areas under the receiver operating characteristic curve; FPD, flat panel detector; I.I.-TV, image intensifier television.

  • * Significantly higher to I.I.-TV system (P < .01).