Pressures, resistances, and flow rates corresponding to the electric circuit analog of patient 2 (Fig 5)
Case | Ps (mm Hg) | Pi (mm Hg) | Po (mm Hg) | Rp (dyne/s/cm5) | Rd (dyne/s/cm5) | Ra (dyne/s/cm5) | Q (mL/s) | ΔP (mm Hg) |
---|---|---|---|---|---|---|---|---|
Pre | 120 | 110 | 100 | 5800 | 57,900 | 5800 | 2.30 | – |
Post | 120 | 112 | 72 | 5800 | 55,000 | 23,200 | 1.90 | 2 |
125 | 117 | 77 | 5800 | 55,000 | 23,200 | 1.98 | 7 | |
130 | 122 | 81 | 5800 | 55,000 | 23,200 | 2.06 | 12 | |
135 | 126 | 86 | 5800 | 55,000 | 23,200 | 2.15 | 16 | |
140 | 130 | 90 | 5800 | 55,000 | 23,200 | 2.22 | 20 | |
145 | 135 | 95 | 5800 | 55,000 | 23,200 | 2.30 | 25 |