Table 1:

Parameters for carotid MR imaging protocols at 3T

ParameterT1WI With or Without Contrast EnhancementT2WITOF ImagingCE-MRA
Acquisition mode2D2D3D3D
Acquisition sequenceFast spin-echoFast spin-echoSpoiled gradient and flow compensationEnhanced gradient recalled-echo
Blood suppression techniqueQuadruple inversion recoveryMultisection double inversion recoverySaturation veinsNone
TE (ms)11523.51.7
TR (ms)8004000235.5
TI (ms)520250n/an/a
Echo-train length1012n/an/a
Excitation flip angle (degrees)90902030
No. of signals acquired121
FOVa160 × 130 mm160 × 120 mm160 × 160 mm150 mm
Matrix size256 × 256256 × 256288 × 256256 × 256
No. of sections18184840
Section thickness (mm)2210.8
Coverage (mm)36364432 mm
Imaging time (min:sec)5:494:324:460:59
  • a The FOV of 140 × 140 cm was also used in some patients.