TABLE 7:

Thresholds beyond which PCT results were significantly different from reference results

Value and RegionTemporal Sampling Interval (seconds)
Group 1 (30 mL)Group 2 (40 mL)Group 3 (50 mL)Group 4 (60 mL)
rCBF
 Healthy hemisphere2334
 Cortical gray matter3345
 White matter2334
 Caudate nucleus3445
 Lenticular nucleus2344
 Ischemic area2334
 Ischemic/healthy ratio2334
rCBV
 Healthy hemisphere3445
 Cortical gray matter3456
 White matter2334
 Caudate nucleus3345
 Lenticular nucleus3345
 Ischemic area2334
 Ischemic/healthy ratio2334
MTT
 Healthy hemisphere3456
 Cortical gray matter4556
 White matter4455
 Caudate nucleus3456
 Lenticular nucleus3446
 Ischemic area3345
 Ischemic/healthy ratio3344
TTP
 Healthy hemisphere3456
 Cortical gray matter4566
 White matter3456
 Caudate nucleus2345
 Lenticular nucleus2445
 Ischemic area2345
 Ischemic/healthy ratio2334
Global2334
  • Note.—Reference results were calculated from the raw acquisition PCT series (temporal sampling interval of 1 second). The final global threshold for each volume was defined as the minimum temporal sampling interval identified as accurate for any of the four parametric PCT results in any region.