Table 1:

Quantitative assessment of metal artifacts before and after reduction

ObjectsHU RangeP ValueNoiseP Value
BeforeAfter% ChangeBeforeAfter% Change
ROI0 (n = 28)247.2 ± 37.5247.3 ± 37.30.04NS36.3 ± 4.836.2 ± 4.70.2NS
Coiling ± stenting (n = 20)2005.6 ± 804.0282.6 ± 55.885.9<.001251.7 ± 94.442.9 ± 10.382.9<.001
    <25 cm (n = 6)1613.7 ± 709.7296 ± 44.781.7.006198.4 ± 83.842.8 ± 7.378.4.007
    25–100 cm (n = 7)1954.9 ± 634.2257.6 ± 23.586.8<.001238.6 ± 49.338.9 ± 5.683.7<.001
    >100 cm (n = 7)2392.3 ± 945.6296 ± 80.787.6.001310.3 ± 113.546.9 ± 15.084.9.001
Liquid embolizer (n = 6)1377.3 ± 646.6428.5 ± 259.068.9.019206.0 ± 86.356.9 ± 21.672.3.007
Shunt (n = 2)1226.5 ± 259.5368.0 ± 76.470.9NS109.8 ± 2.644.0 ± 0.760.013
Overall (ROI1) (n = 28)1815.3 ± 793.7319.9 ± 136.682.3<.001231.7 ± 95.945.9 ± 14.080.2<.001
  • Note:—NS indicates no significance.