Table 1:

Dynamic sequence at the University of Washington is performed on a 3T scanner

ParametersPhilipsaSiemensbGEc
Coil16-Channel neurovascular coil
Parallel imagingSENSEiPATASSET
Sequence3D-T1WI FFE3D-T1WI FISP or 3D-T1WI FLASH3D-T1WI FSPGR
TR/TE for T1 mapping5.2/2.5 ms
MFA for T1 mapping30°, 20°, 15°, 10°, and 2°
TR/TE/FA for dynamic imaging5.2/2.5 ms/5°
FOV212 × 149 mm2
Voxel0.95/0.95/3.00 mm3
Section thickness3 mm
Signal averagingNSA: 1ACQ: 1NEX: 1
Number of sections per dynamic scan/section orientation20/Axial
Temporal resolution3.6 seconds
Total T1 mapping acquisition time26.5 seconds
Total dynamic acquisition time6.10 minutes
Fat saturationNo
Contrast injectionSingle dose of 20-mL gadoteridol (ProHanced) injected at a rate of 5 mL/s through a peripheral arm vein, followed by a 20-mL saline flush with a power injector
  • Note:—FFE indicates fast-field echo; FSPGR, fast-spoiled gradient recalled; MFA, multiple flip angles; SENSE, sensitivity encoding; iPAT, integrated parallel acquisition technique; ASSET, array spatial sensitivity encoding technique; NSA, number of signal averages; ACQ, acquisitions; FA, flip angle.

  • a Phillips Healthcare, Best, the Netherlands.

  • b Siemens, Erlangen, Germany.

  • c GE Healthcare, Milwaukee, Wisconsin.

  • d Bracco Diagnostics, Princeton, New Jersey.