Table 1:

Imaging parameters and conditions for CBCT

10-Second CBCT20-Second CBCT
Exposure parameters
    Tube voltage70 kV70 kV
    Pulse width12.5 ms12.5 ms
    Radiation exposure1.20 μGy/pulse1.20 μGy/pulse
    Binning/focus size/FOV1 × 1/0.4 mm/22 cm
2 × 2/0.6 mm/42 cm
Acquisition parameters
    Angle step0.8°/frame0.4°/frame
    Angle200°200°
    Total frame250500
    Matrix size/binning9602/2 × 2
10242/1 × 1
Injection condition
    Injection rate1.0 mL/s1.0 mL/s
    Concentration of contrast medium14%14%
    X-ray delay time4.5 seconds4.5 seconds
    Injection time of contrast medium14.5 seconds24.5 seconds