Image-quality assessment scores from 35 clinical lumbar spine cases
Scoring Criteria | Mean Score TSE | Mean Score TSE-VFA | P Value of Wilcoxon Testa | Weighted Gwet AC1 |
---|---|---|---|---|
Motion | 4.71 ± 0.59 | 4.83 ± 0.42 | <.001 | 0.88 |
Artifacts | 4.94 ± 0.23 | 4.76 ± 0.46 | <.001 | 0.93 |
Edge sharpness | 4.91 ± 0.33 | 4.79 ± 0.41 | <.001 | 0.93 |
SNR | 4.84 ± 0.40 | 4.36 ± 0.64b | .371b | 0.84 |
Facet joints | 4.83 ± 0.40 | 4.67 ± 0.50 | <.001 | 0.86 |
Endplates | 4.94 ± 0.23 | 4.83 ± 0.38 | <.001 | 0.94 |
Nerve roots | 4.76 ± 0.49 | 4.51 ± 0.58 | <.001 | 0.78 |
Spinal cord | 4.73 ± 0.51 | 4.47 ± 0.61 | <.01 | 0.86 |
Discs | 4.87 ± 0.44 | 4.80 ± 0.44 | <.001 | 0.91 |
↵a The null hypothesis states that the median difference in the image-quality scores between TSE and TSE-VFA is greater than the noninferiority margin Δ, and rejecting the null hypothesis shows noninferiority in performance.