Table 1:

Spatial frequencies of 50%, 10%, and 2% MTF values for UHR acquisitions on both PCD-CT and EID-CT systems with images reconstructed with a sharp kernel (U70)

MTF50%10%2%
PCD-CT with U70 kernel11.2/cm18.4/cm21.1/cm
EID-CT with U70 kernel10.6/cm17.5/cm20.1/cm