Table 1:

Black-blood MR imaging study sequence protocols

2D T1WI TIRM Black-Blood3D T1WI SPACE Black-Blood3D T1WI CS-SPACE Black-Blood
Section orientationAxial and coronalSagittalSagittal
TR, TE, TI (ms), FA, NEX, PAT2000, 11, 1200, 150°, 2, GRAPPA 2710, 11, N/A, N/A, 1, GRAPPA 2. echo train length 157 ms800, 10, N/A, N/A, 1, GRAPPA 2. echo train length 269 ms
FOV230 × 201 mm230 × 230 mm225 × 225 mm
Matrix512 × 410256 × 265448 × 448
Section thickness2 mm (10% gap)0.9 mm0.64 mm
Spatial resolution0.45 × 0.45 × 2 mm0.9 mm (isotropic)0.5 × 0.5 × 0.64 mm
No. of slices15192256
Scan time4 min 6 sec6 min 48 sec6 min 54 sec
  • Note:—FA indicates flip angle; TRIM, turbo inversion recovery magnitude; CS-SSPACE, sampling perfection with application-optimized contrasts by using different flip angle evolutions using compressed sensing; GRAPPA, generalized autocalibrating partially parallel acquisition; PAT, parallel acquisition technique.