Table 1:

Global and peak SAR levels

Duke CenteredElla CenteredCharlie CenteredCharlie −50-mm FHCharlie +50-mm FH
Global SAR levels
 Average SAR for 1 Watt input  power (W kg−1)0.0660.0690.0620.0500.075
 Average SAR per B12 (W kg−1 µT–2)0.4620.4650.2890.4540.474
 Average B1+ in central section  for 1 Watt input power (µT)a0.3790.3850.4660.3330.398
Peak local SAR levels
 Peak local SAR (10-g averaged)  for 1 Watt input power (W kg−1)0.4350.3980.3210.2130.487
 Peak local SAR (10-g averaged)  per B12 (W kg−1µT–2)3.042.631.481.923.08
  • a The power optimization procedure of the MR imaging scanner software calibrates the needed input power to achieve a certain B1 in the subject. This calibration is based on the average B1+ in a central section of the subject (brain in this case).