Abstract
Navigation through a previously deployed and deformed stent is a difficult interventional task. Inadvertent navigation through the struts of a stent can potentially lead to incomplete secondary stent extension and vessel occlusion. Better visualisation of the pathway through the stent can reduce the risks of the procedural complications and reduce the reluctance of the interventionalist to navigate through a previously deployed stent. We describe a technique of visualisation of the pathway navigated by a guidewire through a previously deployed deformed and fractured carotid stent by the use of DynaCT. Three-dimensional reconstruction of the stent/microwire allows excellent visualisation of the correct pathway of the microwire within the stent.
References
Doelken M, Struffert T, Richter G et al (2008) Flat-panel detector volumetric CT for visualization of subarachnoid hemorrhage and ventricles: preliminary results compared to conventional CT. Neuroradiology 50:517–523
White PM, Gilmour JN, Weir NW, Innes B, Sellar RJ (2008) AngioCT in the management of neurointerventional patients: a prospective, consecutive series with associated dosimetry and resolution data. Neuroradiology 50:321–330
Richter G, Engelhorn T, Struffert T et al (2007) Flat panel detector angiographic CT for stent-assisted coil embolization of broad-based cerebral aneurysms. AJNR Am J Neuroradiol 28:1902–1908
Benndorf G, Claus B, Strother CM, Chang L, Klucznik RP (2006) Increased cell opening and prolapse of struts of a neuroform stent in curved vasculature: value of angiographic computed tomography: technical case report. Neurosurgery 58:ONS-E380
Jou LD, Mohamed A, Lee DH, Mawad ME (2007) 3D rotational digital subtraction angiography may underestimate intracranial aneurysms: findings from two basilar aneurysms. AJNR Am J Neuroradiol 28:1690–1692
Bussiere M, Pelz DM, Kalapos P et al (2008) Results using a self-expanding stent alone in the treatment of severe symptomatic carotid bifurcation stenosis. J Neurosurg 109:454–460
Scheinert D, Scheinert S, Sax J et al (2005) Prevalence and clinical impact of stent fractures after femoropopliteal stenting. J Am Coll Cardiol 45:312–315
de Vries JP, Meijer RW, van den Berg JC, Meijer JM, van de Pavoordt ED (2005) Stent fracture after endoluminal repair of a carotid artery pseudoaneurysm. J Endovasc Ther 12:612–615
Valibhoy AR, Mwipatayi BP, Sieunarine K (2007) Fracture of a carotid stent: an unexpected complication. J Vasc Surg 45:603–606
Diehm N, Baum S, Katzen BT, Kovacs M, Benenati J (2008) Fracture of a carotid stent: a word of caution. J Vasc Interv Radiol 19:622–623
Surdell D, Shaibani A, Bendok B, Eskandari MK (2007) Fracture of a nitinol carotid artery stent that caused restenosis. J Vasc Interv Radiol 18:1297–1299
Habib N, Heyden JV (2009) Acute fracture of ACCULINK carotid stent during post dilation. Catheter Cardiovasc Interv. doi:10.1002/ccd.22132
Ling AJ, Mwipatayi P, Gandhi T, Sieunarine K (2008) Stenting for carotid artery stenosis: fractures, proposed etiology and the need for surveillance. J Vasc Surg 47:1220–1226
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Mordasini, P., Al-Senani, F., Gralla, J. et al. The use of flat panel angioCT (DynaCT) for navigation through a deformed and fractured carotid stent. Neuroradiology 52, 629–632 (2010). https://doi.org/10.1007/s00234-009-0556-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00234-009-0556-1