Use of multicoil arrays for separation of signal from multiple slices simultaneously excited

J Magn Reson Imaging. 2001 Feb;13(2):313-7. doi: 10.1002/1522-2586(200102)13:2<313::aid-jmri1045>3.0.co;2-w.

Abstract

Increased acquisition efficiency has been achieved by exciting several slices simultaneously. The mixed data were unfolded to produce separate slices using the spatial encoding information inherent in a multicoil receiver system. Each coil yields a linear combination of signals from all excited slices weighted by the sensitivity of each coil. A matrix inversion provides a solution to unfold these images.

MeSH terms

  • Equipment Design
  • Fourier Analysis
  • Humans
  • Image Processing, Computer-Assisted*
  • Magnetic Resonance Imaging / instrumentation*
  • Phantoms, Imaging
  • Sensitivity and Specificity
  • Time Factors