3.0 Tesla high spatial resolution contrast-enhanced magnetic resonance angiography (CE-MRA) of the pulmonary circulation: initial experience with a 32-channel phased array coil using a high relaxivity contrast agent

Invest Radiol. 2007 Jun;42(6):392-8. doi: 10.1097/01.rli.0000261937.77365.6a.

Abstract

Purpose: To evaluate the technical feasibility of high spatial resolution contrast-enhanced magnetic resonance angiography (CE-MRA) with highly accelerated parallel acquisition at 3.0 T using a 32-channel phased array coil, and a high relaxivity contrast agent.

Materials and methods: Ten adult healthy volunteers (5 men, 5 women, aged 21-66 years) underwent high spatial resolution CE-MRA of the pulmonary circulation. Imaging was performed at 3 T using a 32-channel phase array coil. After intravenous injection of 1 mL of gadobenate dimeglumine (Gd-BOPTA) at 1.5 mL/s, a timing bolus was used to measure the transit time from the arm vein to the main pulmonary artery. Subsequently following intravenous injection of 0.1 mmol/kg of Gd-BOPTA at the same rate, isotropic high spatial resolution data sets (1 x 1 x 1 mm3) CE-MRA of the entire pulmonary circulation were acquired using a fast gradient-recalled echo sequence (TR/TE 3/1.2 milliseconds, FA 18 degrees) and highly accelerated parallel acquisition (GRAPPA x 6) during a 20-second breath hold. The presence of artifact, noise, and image quality of the pulmonary arterial segments were evaluated independently by 2 radiologists. Phantom measurements were performed to assess the signal-to-noise ratio (SNR). Statistical analysis of data was performed by using Wilcoxon rank sum test and 2-sample Student t test. The interobserver variability was tested by kappa coefficient.

Results: All studies were of diagnostic quality as determined by both observers. The pulmonary arteries were routinely identified up to fifth-order branches, with definition in the diagnostic range and excellent interobserver agreement (kappa = 0.84, 95% confidence interval 0.77-0.90). Phantom measurements showed significantly lower SNR (P < 0.01) using GRAPPA (17.3 +/- 18.8) compared with measurements without parallel acquisition (58 +/- 49.4).

Conclusion: The described 3 T CE-MRA protocol in addition to high T1 relaxivity of Gd-BOPTA provides sufficient SNR to support highly accelerated parallel acquisition (GRAPPA x 6), resulting in acquisition of isotopic (1 x 1 x 1 mm3) voxels over the entire pulmonary circulation in 20 seconds.

MeSH terms

  • Adult
  • Aged
  • Artifacts
  • Contrast Media / pharmacokinetics
  • Feasibility Studies
  • Female
  • Humans
  • Image Processing, Computer-Assisted
  • Magnetic Resonance Angiography / methods*
  • Male
  • Meglumine / analogs & derivatives
  • Meglumine / pharmacokinetics
  • Middle Aged
  • Organometallic Compounds / pharmacokinetics
  • Phantoms, Imaging
  • Prospective Studies
  • Pulmonary Circulation*
  • Statistics, Nonparametric

Substances

  • Contrast Media
  • Organometallic Compounds
  • gadobenic acid
  • Meglumine