A case of pedicle screw loosening treated by modified transpedicular screw augmentation with polymethylmethacrylate

J Korean Neurosurg Soc. 2011 Jan;49(1):75-8. doi: 10.3340/jkns.2011.49.1.75. Epub 2011 Jan 31.

Abstract

We report a case of pedicle screw loosening treated by modified transpedicular screw augmentation technique using polymethylmethacrylate (PMMA), which used the anchoring effect of hardened PMMA. A 56-year-old man who had an L3/4/5 fusion operation 3 years ago complained of continuous low back pain after this operation. The computerized tomography showed a radiolucent halo around the pedicle screw at L5. We augmented the L5 pedicle screw with modified pedicle screw augmentation technique using PMMA and performed an L3/4/5 pedicle screw fixation without hook or operation field extension. This modified technique is a kind of transpedicular stiffness augmentation using PMMA for the dead space around the loosed screw. After filling the dead space with 1-2 cc of PMMA, we inserted a small screw. Once the PMMA hardened, we removed the small screw and inserted a thicker screw along the existing screw threading to improve the pedicle screws' pullout strength. At 10 months' follow-up, x-ray showed strong fusion of L3/4/5. The visual analogue scale (VAS) of his back pain was improved from 9 to 5. This modified transpedicular screw augmentation with PMMA using anchoring effect is a simple and effective surgical technique for pedicle screw loosening. However, clinical analyses of long-term follow-up and biomechanical studies are needed.

Keywords: Instrument failure; Osteoporosis; PMMA; Pedicle screw loosening; Pseudoarthrosis; Surgical technique.

Publication types

  • Case Reports