The adverse effects of flow-diverter stent-like devices on the flow pattern of saccular intracranial aneurysm models: computational fluid dynamics study

Acta Neurochir (Wien). 2011 Aug;153(8):1633-40. doi: 10.1007/s00701-011-1055-9. Epub 2011 Jun 8.

Abstract

Background: Stent deployment across the aneurysmal neck has been established as one of the endovascular methods to treat intracranial aneurysms with or without coils.

Objective: The purpose is to study the possible adverse effects of deployment of the new flow-diverter stent-like devices (FD) on the flow characteristics of saccular aneurysm models.

Methods: Numerical simulations of the blood flow patterns in the artificial models of three aneurysms were studied. One model was designed without an FD stent, the second model with one FD stent, and the third model with two stents. Numerical simulation for incompressible laminar blood flow was conducted in the three artificial cerebral aneurysm models by means of computational fluid dynamics.

Results: There was a noticeable increase in the values of the circumferential pressure distributed on the walls of the aneurysm after stent deployment; this led to an increase the tension of the aneurysm surface and was considered to be an adverse effect. This pressure increase was further aggravated by the deployment of another stent. However, there is a beneficial effect of using FD stents, translating into the reduction of the flow velocity inside the aneurysm and wall shear stress at the inflow zone. This reduction decreases further with the deployment of another stent.

Conclusion: Aneurysms become tenser after the deployment of one flow-diverter stent and (more tense still) after after the deployment of another stent. This principle should be kept in mind when choosing which group of aneurysms is the best candidate for such a treatment strategy. This study recommends deploying several FD stents during endovascular procedures until complete arrest of the blood flow occurs during the procedure; otherwise, the aneurysm may become tenser and dangerous if a slow blood flow jet still exists inside it at the end of the procedure.

Publication types

  • Comparative Study

MeSH terms

  • Aneurysm / physiopathology*
  • Aneurysm / therapy*
  • Blood Vessel Prosthesis Implantation / adverse effects*
  • Blood Vessel Prosthesis Implantation / instrumentation
  • Blood Vessel Prosthesis Implantation / standards
  • Cerebral Revascularization / adverse effects*
  • Cerebral Revascularization / instrumentation
  • Cerebral Revascularization / standards
  • Cerebrovascular Circulation / physiology*
  • Computer Simulation*
  • Humans
  • Hydrodynamics*
  • Stents / adverse effects*
  • Stents / standards