Accuracy of MR phase contrast velocity measurements for unsteady flow

J Magn Reson Imaging. 1995 Jul-Aug;5(4):428-31. doi: 10.1002/jmri.1880050410.

Abstract

The accuracy of MR phase contrast (PC) velocity measurements for unsteady flow has been quantitatively assessed. Spatially resolved velocity fields were measured in a long straight tube using a gated PC technique, and the resulting MR PC velocity data were compared with velocities derived from the analytic Womersley solution to the Navier-Stokes equations governing fluid flow. The overall root-mean-square (rms) difference between the measured and analytic velocities was 1.6 cm s-1 for nominally sinusoidal flow waveforms with peak velocities ranging from 51.6 cm s-1 to 59.8 cm s-1. This rms difference corresponded to 7.5% of the mean fluid velocity, which is similar to the cited accuracy of approximately 5% for MR PC velocimetry for steady flows. Linear regression between the PC velocity measurements and the velocities obtained using the analytic expression was highly significant (r2 = 0.997) and yielded a slope of 0.998, close to the expected value of 1. We conclude that the gated MR PC velocity measurements in unsteady flow are accurate.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Blood Flow Velocity
  • Hemorheology
  • Humans
  • Image Processing, Computer-Assisted
  • Linear Models
  • Magnetic Resonance Imaging / methods*
  • Phantoms, Imaging